25
8. Decrease the respiration rate setting on the respiration simulator to 20 breaths per
minute.
a. Verify that the monitor displays the respiration rate of 20 ±3 breaths per minute.
Note: The accuracy of Respiration measurements is ±3 breaths per minute. In the
procedure above, add the tolerance of the simulator to the acceptable range of
readings.
Temperature Operation
1. Press the Power Button to turn on the monitor.
2. Connect the temperature probe (supplied with the temperature simulator) to an
appropriate terminal on the temperature simulator.
3. Connect the temperature probe to the temperature connector on the monitor’s right
panel.
4. Set the temperature simulator as follows:
z Temperature: 37°C (98.0°F)
z Probe type: YSI-400 series Temperature Probes (Probe accuracy: ±0.1°C)
5. After the normal power-up sequence, verify that the temperature reads 37°C ±0.1°C
(98.6°F ±0.2°F if Fahrenheit is selected for the temperature unit).
6. Turn off the monitor.
Note: The accuracy of temperature measurements is ±0.1°C (±0.2°F). In the procedure
above, add the tolerance of the simulator and the probe to the acceptable range of
readings.
CO
2
Operation
1. Connect the monitor to an AC power source.
2. Turn on the monitor by pressing the Power Button.
3. Test #1: Display Accuracy
z For Mainstream
a. Connect the mainstream CO
2
sensor into the CO
2
connector on the monitor’s
right panel.
b. Place a mainstream CO
2
airway adapter into the transducer head.
c. Verify that the “CO
2
- Sensor Warming-up” message displays and wait until
the message clears.
d. Breathe through the CO
2
airway adapter for five slow breaths. Verify that the
waveform displayed rises and falls accordingly.
e. Connect a gas cylinder containing 10% CO
2
to a flow meter and then to the
CO2 airway adapter.
f. Turn the gas ON and verify that the flow through the airway adapter is 200 to
300 ml/min
g. Allow the reading to stabilize for 15 seconds.
h. Verify that the monitor displays 76 ± 2 mmHg (10 ± 0.2%) in CO
2
numerical
area.