5. NIBP
Symptom 5.1: The cuff does not inflate.
Cause or Checkpoint Action Remark
Cuff or Cuff hose is folded. Unfold the cuff or cuff hose.
NIBP tube inside of the monitor is
blocked or kinked.
Check the tube assembly
between NIBP module and hose
fitting.
NIBP module is broken. Replace the NIBP module.
Main board is broken. Replace the Main Board.
6. SpO
2
Symptom 6.1: The SpO
2
sensor does not light.
Cause or Checkpoint Action Remark
The connection between SpO
2
sensor and extension cable is
loose.
Reconnect sensor and extension
cable.
SpO
2
sensor is broken. Replace the SpO
2
sensor.
SpO
2
module is broken. Replace the SpO
2
module.
Main Board is broken. Replace the Main Board.
Symptom 6.2: No display on the numeric window though sensor is fine.
Cause or Checkpoint Action Remark
SpO
2
module is broken. Replace the SpO
2
module.
7. Temperature
Symptom 7.1: Temperature value does not display.
Cause or Checkpoint Action Remark
Sensor is not properly located. Check the connection of
Temperature sensor to the skin.
Temperature module is broken. Replace the Temperature
module.
Main Board is broken. Replace the Main Board.
8. Respiration
Symptom 8.1: Respiration value is inaccurate.
Cause or Checkpoint Action Remark
Gain is too low. Select appropriate gain from the
Respiration menu.
Electrodes are not attached tightly. Check the electrodes.