Heart rate meter accuracy
and response to irregular
rhythm
Provides correct heart rates, as follows:
Ventricular bigeminy: 76 to 80 bpm
Slow alternating ventricular bigeminy: 55 to 60 bpm
Rapid alternating ventricular bigeminy: 116 to 117
bpm
Bidirectional systoles: 86 to 91 bpm
Respiration, leads-off
sensing, and active noise
suppression
Respiration frequency: 61 to 62 kHz
Lead-off sensing current:
amp: less than 0.02uA
common: less than 0.078uA
Time to alarm for cardiac
standstill
Average: 4.0 sec
Time to alarm for low heart
rate
Average: 5.5 sec
Time to alarm for high heart
rate
Average: 5.6 sec
Time to alarm for
tachycardia
Ventricular tachycardia waveform
Amplitude 1mV, Heart rate 206bpm: 16.48 sec
One-half waveform (0.5mV): 11.64 sec
Twice waveform (2mV): 5.81 sec
Amplitude 2mV, Heart rate 195bpm: 8.74 sec
One-half waveform (1mV): 12.03 sec
Twice waveform (2mV): 1.75 sec