41
Factory Defaults
Parameter Ranges/Selections
Adult Pediatric Neonatal
NIBP DIA Upper Alarm
Limits
15 to 250 mmHg (Adult/Pedi)
2.0 to 33.3 kPa (Adult/Pedi)
25 to 90 mmHg (Neo)
3.3 to 12.0 kPa (Neo)
(5 mmHg / 0.6 or 0.7 kPa steps)
90 mmHg
12.0 kPa
70 mmHg
9.3 kPa
60 mmHg
8.0 kPa
NIBP DIA Lower Alarm
Limits
10 to 245 mmHg (Adult/Pedi)
1.3 to 32.6 kPa (Adult/Pedi)
20 ~85 mmHg (Neo)
2.6 to 11.3 kPa (Neo)
(5 mmHg / 0.6 or 0.7 kPa steps)
50 mmHg
6.6 kPa
40 mmHg
5.3 kPa
20 mmHg
2.6 kPa
NIBP MAP Upper Alarm
Limits
25 to 260 mmHg (Adult/Pedi)
3.3 to 34.6 kPa (Adult/Pedi)
35 to 110 mmHg (Neo)
4.6 to 14.6 kPa (Neo)
(5 mmHg / 0.6 or 0.7 kPa steps)
110 mmHg
14.6 kPa
90 mmHg
12.0 kPa
70 mmHg
9.3 kPa
NIBP MAP Lower Alarm
Limits
20 to 255 mmHg (Adult/Pedi)
2.6 to 34.0 kPa (Adult/Pedi)
30 to 105 mmHg (Neo)
4.0 to 14.0 kPa (Neo)
(5 mmHg / 0.6 or 0.7 kPa steps)
60 mmHg
8.0 kPa
50 mmHg
6.6 kPa
30 mmHg
4.0 kPa
SpO
2
C-Lock On, Off Off Off Off
%SpO
2
Alarm Inhibition On, Off Off Off Off
%SpO
2
Upper Alarm Limits 21 to 100 % (Adult/Pedi/Neo) (1 % steps) 100 % 100 % 100 %
%SpO
2
Lower Alarm Limits 20 to 99 % (Adult/Pedi/Neo) (1 % steps) 90 % 90 % 85 %
Respiration
Respiration On, Off On On On
Respiration source AUTO, AW, IM AUTO AUTO AUTO
Respiration Alarm
Inhibition
On, Off Off Off Off
RR Upper Alarm Limits 4 to 120 BPM (1 BPM steps) 30 BPM 30 BPM 100 BPM
RR Lower Alarm Limits 3 to 119 BPM (1 BPM steps) 8 BPM 8 BPM 30 BPM
Temperature
Temp Unit °C, °F °C °C °C
Temp Alarm Inhibition On, Off Off Off Off
Temp Upper Alarm Limits 0.1 to 50.0 °C (Adult/Pedi/Neo)
(0.1º C steps)
32.2 to 122.0 °F (Adult/Pedi/Neo)
(0.1°F or 0.2°F steps)
39.0 °C
(102.2 °F)
39.0 °C
(102.2 °F)
39.0 °C
(102.2 º F)
Temp Lower Alarm Limits 0.0 to 49.9 °C (Adult/Pedi/Neo)
(0.1º C steps)
32.0 to 122.8 °F (Adult/Pedi/Neo)
(0.1°F or 0.2°F steps)
36.0 °F
(96.8 °F)
36.0 °F
(96.8 °F)
36.0 °F
(96.8 °F)
EtCO
2
EtCO
2
Unit mmHg, kPa, % mmHg mmHg mmHg
APNEA Alarm Off, 15, 20, 25, 30, 35, 40 sec 20 sec 20 sec 20 sec
Zero Calibration Yes, No No No No
EtCO
2
Alarm Inhibition On, Off Off Off Off
InCO
2
Alarm linhibition On, Off Off Off Off